Medico Leadlok Solid Gel Electrodes

All Medico Leadlok solid gel electrodes are tested according to the voluntary Association for the Advancement of Medical Instrumentation (AAMI) standard requirements for electrical performance for disposable ECG electrodes and exceed those performance standards.
 

Stealth Solid Gel Electrodes


The following products are injected with our proprietary Stealth Solid Gel:

MSGLT - 01
MSGLT - 03
MSGLT - 04
MSGLT - 05
MSGLT - 06
MSGLT - 08
MSGLT - 09
MSGLT - 12

Test Results:


 

Medico Leadlok Typical
AAMI Limits

 DC Offset Voltage (mV)

0.12
 Not more than 100

 AC Impedance @ 10Hz (OHMS)

60
 Not more than 2000

 DC Offset @ 5 sec after capacitor discharge (mV)

6.5
 Not more than 100

 Recovery Slope @ 30 second interval (mV/sec)

0.0
 Not more than 1.0

 AC Impedance after defib (OHMS)

46
 Not more than 3000

HOME       |         PRODUCTS       |         TEST RESULT         |         DOWNLOADS         |         CONTACT US

 

                 Copyright © 2007
Medico Electrodes International ECG Electrodes Manufacturers