Medico Leadlok Liquid Gel Electrodes
 
Potassium Chloride Liquid Gel

The following products incorporate our proprietary Potassium Chloride Liquid Gel and is tested according to the voluntary Association of the Advancement of Medical Instrumentation (AAMI) standard requirements for electrical performance for disposable ECG electrodes:
 
 
MLGLT-01
MLGLT-02
MLGLT-03
MLGST-06

Test Results:


 

Medico Leadlok Typical
AAMI Limits
 DC Offset Voltage (mV)
0.15
 Not more than 100

 AC Impedance @ 10Hz (OHMS)

51
 Not more than 2000

 DC Offset @ 5 sec after capacitor discharge (mV)

0.7
 Not more than 100

 Recovery Slope @ 30 second interval (mV/sec)

0.0
 Not more than 1.0

 AC Impedance after defib (OHMS)

49
 Not more than 3000
 
 